Indigenous Development of a Clinically Safe, Low Cost Microcontroller-Based Blood Warmer for Resource Limited Healthcare

Authors

  • Yasir Ali Shah Department of Mechanical Engineering, University of Engineering & Technology, Mardan, Mardan-23200, Pakistan Author
  • Malik Muhammad Aqib Khan Department of Mechanical Engineering, University of Engineering & Technology, Mardan, Mardan-23200, Pakistan Author
  • Muhammad Awais Khan Department of Mechanical Engineering, University of Engineering & Technology, Mardan, Mardan-23200, Pakistan Author
  • Malik Sarmad Zahid Department of Mechanical Engineering, University of Engineering & Technology, Mardan, Mardan-23200, Pakistan Author
  • Mushaf Ur Rehman Department of Mechanical Engineering, University of Engineering & Technology, Mardan, Mardan-23200, Pakistan Author

DOI:

https://doi.org/10.54536/ajmhc.v1i1.6613

Keywords:

Blood Warmer, Blood Transfusion, Cost-Effective, Hemolysis, Resource-Limited Healthcare, Thermal Performance

Abstract

Blood transfusion is a critical medical procedure in trauma care, surgery, and obstetrics. However, transfusion cold-stored blood without proper warming can induce hypothermia, cardiac arrhythmias, and increase mortality rates, particularly in resource-limited healthcare environments such as rural Pakistan. In the present study, we design and fabricate an indigenous, low-cost, microcontroller-based blood warming system optimized for deployment in resource-limited medical settings. The device incorporates Positive Temperature Coefficient (PTC) ceramic heating elements, a servo-actuated shaking mechanism to ensure uniform heat distribution, thermocol insulation for thermal retention, and an MLX90614 infrared sensor for precise, non-contact temperature monitoring. A simple on-off control algorithm maintains stable and reliable temperature regulation at the clinical threshold of 37 °C, with an integrated buzzer and Liquid Crystal Display (LCD) interface providing real-time operational feedback and safety alerts. Bench-top in vitro tests demonstrated that the prototype effectively elevated a 500 mL blood bag from 4-6 °C to 37 °C within 12-14 minutes, achieving high accuracy (±0.4 °C) and uniform warming without inducing hemolysis, as confirm by plasma free hemoglobin analysis. Comparative evaluation against a laboratory water bath highlighted the device’s faster and more controlled heating profile. Cost analysis revealed a fabrication cost of approximately PKR 29,000 (˂100 USD), substantially lower than commercial alternatives. Thus, the developed system offers an affordable, safe, power efficient (~300 W), and ease of local fabrication, presenting a clinically viable and scalable solution for safe blood transfusion in emergency and resource-limited healthcare settings.

Downloads

Download data is not yet available.

References

Brohi, K., Gruen, R. L., & Holcomb, J. B. (2019). Why are bleeding trauma patients still dying? Intensive Care Medicine, 45(5), 709–711. https://doi.org/10.1007/s00134-019-05560-x

C. Smith, R. Desai, V. Glorioso, A. C.-… of C. Anesthesia, and undefined 1998, “Preventing hypothermia: convective and intravenous fluid warming versus convective warming alone,” Elsevier,

C. S.-S. in anesthesia, perioperative medicine and, and undefined 2001, “Principles of fluid warming in trauma,” Elsevier, Accessed: Aug. 27, 2025

Cherry, M. S., Hodgson, G. H., & Nottebrock, H. (1981). Comparison of two in-line blood warmers. Canadian Anaesthetists’ Society Journal, 28(2), 180–181. https://doi.org/10.1007/BF03007266

D. S.-A. C. of N. America and undefined 1994, “Consequences and treatment of perioperative hypothermia,” Elsevier, https://www.sciencedirect.com/science/article/pii/S0889853721006866

E. Poppa, D. Yona, Y. Katz, M. B.-J. of clinical anesthesia, and undefined 2009, “Warmflo® warming system overheats fluids in simulated clinical conditions,” Elsevier, https://www.sciencedirect.com/science/article/pii/S095281800900169X

Goodnough, L., Levy, J., & Murphy, M. F. (2013). Concepts of blood transfusion in adults. The Lancet. https://www.thelancet.com

Henker, R., Bernardo, L., & [Additional authors]. (1995). Evaluation of four methods of warming intravenous fluids. Journal of Emergency Medicine. Elsevier.

H. Hendolin, E. L.-A. of clinical research, and undefined 1982, “Skin and central temperatures during continuous epidural analgesia and general anaesthesia in patients subjected to open prostatectomy.,” europepmc.org,

Hijji, B., Oweis, A., & Dabbour, R. S. (2012). Measuring knowledge of blood transfusion: A survey of Jordanian nurses. American International Journal of Contemporary Research. Retrieved August 27, 2025, from https://www.academia.edu/download/95371533/10.pdf

Iserson, K. V., & Huestis, D. W. (1991). Blood warming: Current applications and techniques. Transfusion, 31(6), 558–571. https://doi.org/10.1046/j.1537-2995.1991.31691306256.x

Iserson, K. V.-B. M. J.-I. (2014). Rapid admixture blood warming: Fast, safe & inexpensive. BMH Medical Journal, 1(3), 40–46.

Joarder, M. K., Ferdous, A. S. M. A., & Ali, S. N. (2022). Development of a low-cost prototype of blood warmer and rapid infuser. In 12th International Conference on Electrical and Computer Engineering (ICECE 2022) (pp. 401–404). IEEE. https://doi.org/10.1109/ICECE57408.2022.10088824

L. Uhl, D. Pacini, M. K.- Anesthesiology, and undefined 1992, “A comparative study of blood warmer performance.,” europepmc.orgL Uhl, D Pacini, MS KruskallAnesthesiology, 1992•europepmc.org,

Leaman, P., & [Additional author if any]. (1985). Microwave warming of resuscitation fluids. Annals of Emergency Medicine. Retrieved August 27, 2025, from https://www.sciencedirect.com/science/article/pii/S0196064485806375

Mommsen, P., Andruszkow, H., Frömke, C., Zeckey, C., Wagner, U., Van Griensven, M., & Frink, M. (2013). Effects of accidental hypothermia on posttraumatic complications and outcome in multiple trauma patients. Injury, 44(1), 86. https://doi.org/10.1016/j.injury.2011.10.013

M. Vaughan, R. Vaughan, R. C.-A. & Analgesia, and undefined 1981, “Postoperative hypothermia in adults: relationship of age, anesthesia, and shivering to rewarming,” journals.lww.com,

Muth, C. M., Mainzer, B., & Peters, J. (1996). The use of countercurrent heat exchangers diminishes accidental hypothermia during abdominal aortic aneurysm surgery. Acta Anaesthesiologica Scandinavica, 40(10), 1197–1202. https://doi.org/10.1111/j.1399-6576.1996.tb05550.x

S. Insler, D. S.-A. C. of, and undefined 2006, “Perioperative thermoregulation and temperature monitoring,” anesthesiology.theclinics.

R. L.-B. P. & R. C. Anaesthesiology and undefined 2003, “Monitoring and thermal management,” Elsevier, Accessed: Aug. 27, 2025.

Platts-Mills, T. F., Stendell, E., Lewin, M. R., Moya, M. N., Dhah, K., Stroh, G., & Shalit, M. (2007). An experimental study of warming intravenous fluid in a cold environment. Wilderness & Environmental Medicine, 18(3), 177–185. https://doi.org/10.1580/06-WEME-OR-051R1.1

Roxby, D., Sobieraj-Teague, M., Von Wielligh, J., Sinha, R., Kuss, B., Smith, A. L., & McEwen, M. (2020). Warming blood prior to transfusion using latent heat. Emergency Medicine Australasia, 32(4), 604–610. https://doi.org/10.1111/1742-6723.13471

Suksantisakul, P., Maneerat, N., & [Additional authors]. (2014). Blood warmer using Peltier devices. International Journal of Bioscience, Biochemistry and Bioinformatics. Retrieved September 3, 2025, from https://www.academia.edu/download/100232875/ijbsbt.2014.6.3.pdf

Turner, M., Hodzovic, I., & Mapleson, W. W. (2006). Simulated clinical evaluation of four fluid warming devices. Anaesthesia, 61(6), 571–575. https://doi.org/10.1111/j.1365-2044.2006.04589.x

Downloads

Published

2026-04-12

How to Cite

Shah, Y. A. ., Khan, M. M. A. ., Khan, M. A. ., Zahid, M. S. ., & Rehman, M. U. . (2026). Indigenous Development of a Clinically Safe, Low Cost Microcontroller-Based Blood Warmer for Resource Limited Healthcare. American Journal of Medicine and Health Care, 1(1), 22-29. https://doi.org/10.54536/ajmhc.v1i1.6613